Subtrochanteric shortening in total hip arthroplasty: biomechanical comparison of four techniques.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 24095585.
- Also identified by DOI 10.1016/j.arth.2013.09.004.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Safe reduction of the femoral head into the true acetabulum requires a certain amount of femoral shortening in patients with high dislocation of the hip. In subtrochanteric shortening applications, to reduce complications it is necessary to maintain a stable fixation at the osteotomy line. The purpose of this study is to investigate frequently used methods from a biomechanical point of view. Four osteotomy groups were created with composite femurs to investigate subtrochanteric osteotomies; transverse, oblique, z-subtrochanteric and double Chevron. All loading tests were carried out with two implant types both with and without strut graft and cable fixation. No single inherent feature increasing the stability of the investigated osteotomy types was found. Additionally graft application did not have a significant contribution to stability.