Predicting wear and blood metal ion levels in metal-on-metal hip resurfacing.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 24115200.
- Also identified by DOI 10.1002/jor.22459.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Suboptimal component position and design are thought to lead to edge wear and raised blood metal ion levels in metal-on-metal hip resurfacing (MOM-HR). These factors are thought to influence the "contact patch to rim distance" (CPRD), and calculation of this distance may improve prediction of wear and blood metal ion levels. We measured blood cobalt and chromium ion levels and the wear rates of the bearing surfaces in 165 MOM-HR retrieval cases. We then determined the contribution and effect sizes of cup inclination and version angles, component size and design, and CPRD (calculated from case-specific data) on blood metal ion levels and component wear rates. Acetabular orientation explained between 16.3% and 28.5% of the variation in wear rates and metal ion levels, whereas component size and design explained between 7.3% and 21.8% of the variability. In comparison, CPRD explained up to 67.7% of the variability, significantly greater than any other variable (all p < 0.0001). CPRD is a good predictor of wear and improves our understanding of wear performance and the mechanisms leading to edge loading.
Medical subject headings
- Arthroplasty, Replacement, Hip
- Chromium
- Cobalt
- Equipment Failure Analysis
- Hip Prosthesis
- Prosthesis Design
Anatomy
- hip