Results of titanium locking plate and stainless steel cerclage wire combination in femoral fractures.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 24133304.
- Also identified by DOI 10.4103/0019-5413.118200 and PMC identifier 3796917.
- Licence recorded as CC BY-NC-SA.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Some in vitro studies warn combining different metals in orthopedic surgery. The aim of this study is to determine the impact of combining titanium and stainless steel on bone healing and the clinical course of patients undergoing internal fixation of femoral fractures. 69 patients with femoral fractures had polyaxial locking plate osteosynthesis. The locking plate was made of a titanium alloy. Two different cohorts were defined: (a) sole plating and (b) additional stainless steel cerclage wiring. Postoperative radiographs and clinical followup were performed at 6 weeks, 3 months and 12 months. Cohorts A and B had 36 and 33 patients, respectively. Patient demographics and comorbidities were similar in both groups. In two cases in cohort A, surgical revision was necessary. No complication could be attributed to the combination of titanium and stainless steel. The combination of stainless steel cerclage wires and titanium plates does not compromise fracture healing or the postoperative clinical course.
Anatomy
- femur