Model-based optoacoustic image reconstruction of large three-dimensional tomographic datasets acquired with an array of directional detectors.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 24144658.
- Also identified by DOI 10.1109/TMI.2013.2286546.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Image quality in 3-D optoacoustic (photoacoustic) tomography is greatly influenced by both the measurement system, in particular the number and spatial arrangement of ultrasound sensors, and the ability to account for the spatio-temporal response of the sensor element(s) in the reconstruction algorithm. Herein we present a reconstruction procedure based on the inversion of a time-domain forward model incorporating the spatial impulse response due to the shape of the transducer, which is subsequently applied in a tomographic system based on a translation-rotation scan of a linear detector array. The proposed method was also adapted to cope with the data-intensive requirements of high-resolution volumetric optoacoustic imaging. The processing of 2 · 10 (4) individual signals resulted in well-resolved images of both ~ 200 μm absorbers in phantoms and complex vascular structures in biological tissue. The results reported herein demonstrate that the introduced model-based methodology exhibits a better contrast and resolution than standard back-projection and model-based algorithms that assume point detectors. Moreover, the capability of handling large datasets anticipates that model-based methods incorporating the sensor properties can become standard practice in volumetric opto acoustic image formation.
Medical subject headings
- Imaging, Three-Dimensional
- Photoacoustic Techniques
- Tomography, X-Ray Computed