CT scan parameters and radiation dose: practical advice for radiologists.
Level V
Where this comes from
- Record sourced from PubMed, PMID 24183553.
- Also identified by DOI 10.1016/j.jacr.2013.05.032.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Although there has been increasing recognition of the importance of reducing radiation dose when performing multidetector CT examinations, the increasing complexity of CT scanner technology, as well as confusion about the importance of many different CT scan parameters, has served as an impediment to radiologists seeking to create lower dose protocols. The authors seek to guide radiologists through the manipulation of 8 fundamental CT scan parameters that can be altered or optimized to reduce patient radiation dose, including detector configuration, tube current, tube potential, reconstruction algorithm, patient positioning, scan range, reconstructed slice thickness, and pitch. Although there is always an inherent trade-off between image quality or noise and patient radiation dose, in many cases, a reasoned manipulation of these 8 parameters can allow the safer imaging of patients (with lower dose) while preserving diagnostic image quality.
Medical subject headings
- Algorithms
- Patient Safety
- Radiation Dosage
- Radiation Injuries
- Radiation Protection
- Radiographic Image Enhancement
- Tomography, X-Ray Computed