A laboratory study of the effects of processing blood through a cell salvage device and leucocyte depletion filter on levels of pro-inflammatory cytokines and bradykinin.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 24219252.
- Also identified by DOI 10.1111/anae.12420.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We investigated changes in concentrations of interleukin-1β, interleukin-6, tumour necrosis factor-α and bradykinin in blood during passage through a cell salvage device and a leucocyte depletion filter, with or without application of subatmospheric pressure across the filter. Blood samples from 19 healthy women undergoing scheduled caesarean section showed concentrations of cytokines and bradykinin in blood filtered under gravity flow that were equal to or significantly lower than those of pre-operative venous blood samples. They were also significantly lower than that in postoperative orthopaedic shed blood, which is commonly reinfused after orthopaedic surgery. A minority of samples taken from blood that had been filtered using subatmospheric pressure showed raised interleukin-6 concentrations. We suggest that use of a leucocyte depletion filter for cell-salvaged blood with gravity flow is likely to be safe with regard to concentrations of cytokines and bradykinin. However, this may not hold true for the filter used with subatmospheric pressure. If transfusion of salvaged blood using a leucocyte depletion filter seems to induce hypotension, elevation of interleukin-6 should be suspected.
Medical subject headings
- Blood Transfusion, Autologous
- Bradykinin
- Cytokines
- Filtration
- Leukocyte Reduction Procedures
- Operative Blood Salvage