Comparison of acetabular shell position using patient specific instruments vs. standard surgical instruments: a randomized clinical trial.
rct · Level II
Where this comes from
- Record sourced from PubMed, PMID 24231437.
- Also identified by DOI 10.1016/j.arth.2013.10.006.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Total hip arthroplasty (THA) survivorship relies largely upon appropriate acetabular cup placement. The purpose of this prospective randomized controlled trial was to determine whether the use of a preoperative 3D planning software in combination with patient specific instrumentation (PSI) results in improved cup placement compared with traditional techniques. Thirty-six THA patients were randomized into standard (STD) or PSI technique. Standard approach was completed using traditional techniques, while PSI cases were planned and customized surgical instruments were manufactured. Postoperative CT scans were used to compare planned to actual results. Differences found between planned and actual anteversion were -0.2° ± 6.9° (PSI) and -6.9°±8.9° (STD) (P = 0.018). Use of 3D preoperative planning along with PSIs resulted in significantly greater anteversion accuracy than traditional planning and instrumentation.