Poor long-term stability of the Corvita abdominal stentgraft.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 24231455.
- Also identified by DOI 10.1016/j.ejvs.2013.10.010.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The endovascular treatment of abdominal aortic aneurysms was introduced in the early 1990s, with different generations of devices using various options for either the stent skeleton or the membrane. Corvita generated one of these devices using braided stainless steel and a porous spun polycarbonate urethane membrane. In this report, we describe a case involving Corvita stentgraft explantation for complete aneurysm reperfusion after 13 years, demonstrating major degradation of the polyurethane membrane.