Initial stability of press-fit acetabular components under rotational forces.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 24238907.
- Also identified by DOI 10.1016/j.arth.2013.10.009.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The primary goal of this study was to determine the initial press-fit stability in acetabular components without screw fixation. Mechanical testing was performed with the implantation of press-fit acetabular components in cadaveric specimens. No significant difference was found in load to failure testing between 1 and 2 mm of under-reaming. However, there was significant variability in bending forces required to create 150 μm of micromotion ranging from 49.3 N to 214.4 N. This study shows that cups implanted in a press-fit fashion, which are felt to be clinically stable, have high degrees of variability in resisting load and may be at risk for loosening. There is a need for more objective intra-operative techniques to test cup stability.