Miniaturized passive hydrogel check valve for hydrocephalus treatment.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 24239966.
- Also identified by DOI 10.1109/TBME.2013.2290291.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Improvements in cerebrospinal fluid (CSF) draining techniques for treatment of hydrocephalus are urgently sought after to substitute for current CSF shunts that are plagued by high failure rates. The passive check valve aims to restore near natural CSF draining operations while mitigating possible failure mechanisms caused by finite leakage or low resilience that frequently constrain practical implementation of miniaturized valves. A simple hydrogel diaphragm structures core passive valve operations and enforce valve sealing properties to substantially lower reverse flow leakage. Experimental measurements demonstrate realization of targeted cracking pressures (PT ≈ 20-110 mmH2O) and operation at -800 <; ΔP <; 600 mmH2O without observable degradation or leakage.
Medical subject headings
- Hydrocephalus
- Hydrogel, Polyethylene Glycol Dimethacrylate
- Intracranial Pressure
- Microtechnology
- Monitoring, Physiologic