Long-term hip simulator testing of the artificial hip joint bearing surface grafted with biocompatible phospholipid polymer.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 24249706.
- Also identified by DOI 10.1002/jor.22526.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To prevent periprosthetic osteolysis and subsequent aseptic loosening of artificial hip joints, we recently developed a novel acetabular highly cross-linked polyethylene (CLPE) liner with graft polymerization of 2-methacryloyloxyethyl phosphorylcholine (MPC) on its surface. We investigated the wear resistance of the poly(MPC) (PMPC)-grafted CLPE liner during 20 million cycles in a hip joint simulator. We extended the simulator test of one liner to 70 million cycles to investigate the long-term durability of the grafting. Gravimetric, surface, and wear particle analyses revealed that PMPC grafting onto the CLPE liner surface markedly decreased the production of wear particles and showed that the effect of PMPC grafting was maintained through 70 million cycles. We believe that PMPC grafting can significantly improve the wear resistance of artificial hip joints.
Medical subject headings
- Hip Prosthesis
- Methacrylates
- Phosphorylcholine
- Polyethylene
Anatomy
- hip