The development of a dynamic, six-axis spine simulator.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 24321131.
- Also identified by DOI 10.1016/j.spinee.2013.11.045.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Although a great deal of research has been completed to characterize the stiffness of spinal specimens, there remains a limited understanding of the spine in 6 df and there is a lack of data from dynamic testing in six axes. This study details the development and validation of a dynamic six-axis spine simulator. Biomechanical study. A synthetic spinal specimen was used for the purpose of tuning the simulator, completing positional accuracy tests, and measuring frequency response under physiological conditions. The spine simulator was used to complete stiffness matrix tests of an L3-L4 lumbar porcine functional spinal unit. Five testing frequencies were used, ranging from quasistatic (0.00575 Hz) to dynamic (0.5 Hz). Tests were performed without an axial preload and with an axial preload of 500 N. The validation tests demonstrated that the simulator is capable of producing accurate positioning under loading at frequencies up to 0.5 Hz using both sine and triangle waveforms. The porcine stiffness matrix tests demonstrated that the stiffness matrix is not symmetrical about the principal stiffness diagonal. It was also shown that while an increase in test frequency generally increased the principal stiffness terms, axial preload had a much greater effect. The spine simulator is capable of characterizing the dynamic biomechanics of the spine in six axes and provides a means to better understand the complex behavior of the spine under physiological conditions.
Medical subject headings
- Lumbar Vertebrae
Anatomy
- lumbar spine