Computational fluid dynamics in patients with continuous-flow left ventricular assist device support show hemodynamic alterations in the ascending aorta.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 24345553.
- Also identified by DOI 10.1016/j.jtcvs.2013.09.069.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Increased use of continuous-flow left ventricular assist devices for long-term mechanical support necessitates a better understanding of hemodynamic changes in the native heart and the ascending aorta. By using patient-specific computational models, correlations of potentially adverse hemodynamic conditions with the orientation of the left ventricular assist device outflow graft and their relationship with aortic insufficiency and ischemic events were investigated. Computed hemodynamic parameters, including wall shear stress, pressure in the ascending aorta, and dissipation of turbulent energy, were correlated with the orientation of the left ventricular assist device graft outflow in 5 patients (4 with the HeartMate II device [Thoratec Corp, Pleasanton, Calif] and 1 with the HeartWare Ventricular Assist Device [HeartWare Inc, Framingham, Mass]; 3 patients experienced moderate aortic insufficiency, and 2 patients experienced ischemic events). Hemodynamic conditions for aortic insufficiency and ischemic events were differentiated by linear discriminant analysis. Positive correlations between left ventricular assist device outflow graft orientation and wall shear stress, pressure, and turbulent energy dissipation in the ascending aorta were found (R(2) > 0.68). Linear discriminant analysis indicated a relationship of the velocity magnitude of retrograde flow toward the aortic root with aortic insufficiency and of the turbulent energy and wall shear stress with ischemic events. Computational fluid dynamic simulations using clinical image data indicate altered hemodynamic conditions after left ventricular assist device implantation. Consequently, the left ventricular assist device outflow graft should be placed so the jet of blood is aimed toward the lumen of the aortic arch to avoid turbulences that will increase wall shear stress and retrograde pressure of the aortic root. Further investigations are warranted to confirm these findings in a larger patient cohort.
Medical subject headings
- Aorta
- Aortic Valve Insufficiency
- Computer Simulation
- Heart-Assist Devices
- Hemodynamics
- Hydrodynamics
- Models, Cardiovascular