Time-delay when updating infusion rates in the Graseby 3400 pump results in reduced drug delivery.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 24356164.
- Also identified by DOI 10.1213/01.ane.0000438349.21617.2f.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Infusion pumps are commonly used for infusion of drugs for physiologic control, and infusion rate has been demonstrated to affect the parameters of pharmacokinetic models. In attempting to develop a model that explained this behavior, we examined the behavior of the Graseby 3400 syringe pump under a range of flow conditions and with variations in syringe characteristics. Two issues were identified: start-up loss (the difference between actual and ideal delivery on initial infusion) and update loss (the difference between actual and ideal delivery when transmitting a command to change infusion rate). Under worst-case conditions, this may result in a 20-second period of zero delivery during start-up, and when updating infusion rates once per second, zero output. These effects are influenced by syringe characteristics and vary sufficiently as to make it impossible to isolate this effect from the pharmacokinetic process being controlled. The implications of this for previous published results and clinical application of target-controlled infusions are discussed.