Multicenter harmonization of 89Zr PET/CT performance.
other
Where this comes from
- Record sourced from PubMed, PMID 24357686.
- Also identified by DOI 10.2967/jnumed.113.130112.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
This study investigated the feasibility of quantitative accuracy and harmonized image quality in (89)Zr-PET/CT multicenter studies. Five PET/CT scanners from 3 vendors were included. (89)Zr activity was measured in a central dose calibrator before delivery. Local activity assays were based on volume as well as on the local dose calibrator. Accuracy and image noise were determined from a cross calibration experiment. Image quality was assessed from recovery coefficients derived from different volume-of-interest (VOI) methods (VOI A 50%, based on a 3-dimensional isocontour at 50% of the maximum voxel value with local background correction; VOI Max, based on the voxel with the highest uptake; and VOI 3Dpeak, based on a spheric VOI of 1.2-cm diameter positioned so as to maximize the enclosed average). PET images were analyzed before and after postreconstruction smoothing, applied to match image noise. PET/CT accuracy and image noise ranged from -3% to 10% and from 13% to 22%, respectively. VOI 3Dpeak produced the most reproducible recovery coefficients. After calibration of the local dose calibrator to the central dose calibrator, differences between the local activity assays were within 6%. This study showed that quantitative accuracy and harmonized image quality can be reached in (89)Zr PET/CT multicenter studies.
Medical subject headings
- Algorithms
- Calibration
- Humans
- Image Processing, Computer-Assisted
- Image Processing, Computer-Assisted/methods
- Multicenter Studies as Topic
- Multimodal Imaging
- Multimodal Imaging/methods
- Phantoms, Imaging
- Positron-Emission Tomography
- Positron-Emission Tomography/methods
- Radioisotopes
- Radioisotopes/pharmacology
- Reproducibility of Results
- Tomography, X-Ray Computed
- Tomography, X-Ray Computed/methods
- Zirconium
- Zirconium/pharmacology