Multicenter harmonization of 89Zr PET/CT performance.

Makris, Nikolaos E; Boellaard, Ronald; Visser, Eric P; de Jong, Johan R; Vanderlinden, Bruno; Wierts, Roel; van der Veen, Berlinda J; Greuter, Henri J N M et al. · J Nucl Med · 2014

other

Where this comes from

Abstract

This study investigated the feasibility of quantitative accuracy and harmonized image quality in (89)Zr-PET/CT multicenter studies. Five PET/CT scanners from 3 vendors were included. (89)Zr activity was measured in a central dose calibrator before delivery. Local activity assays were based on volume as well as on the local dose calibrator. Accuracy and image noise were determined from a cross calibration experiment. Image quality was assessed from recovery coefficients derived from different volume-of-interest (VOI) methods (VOI A 50%, based on a 3-dimensional isocontour at 50% of the maximum voxel value with local background correction; VOI Max, based on the voxel with the highest uptake; and VOI 3Dpeak, based on a spheric VOI of 1.2-cm diameter positioned so as to maximize the enclosed average). PET images were analyzed before and after postreconstruction smoothing, applied to match image noise. PET/CT accuracy and image noise ranged from -3% to 10% and from 13% to 22%, respectively. VOI 3Dpeak produced the most reproducible recovery coefficients. After calibration of the local dose calibrator to the central dose calibrator, differences between the local activity assays were within 6%. This study showed that quantitative accuracy and harmonized image quality can be reached in (89)Zr PET/CT multicenter studies.