Microneedle-based transdermal sensor for on-chip potentiometric determination of K(+).

Miller, Philip R; Xiao, Xiaoyin; Brener, Igal; Burckel, D Bruce; Narayan, Roger; Polsky, Ronen · Adv Healthc Mater · 2014

basic_science · Level V

Where this comes from

Abstract

The determination of electrolytes is invaluable for point of care diagnostic applications. An ion selective transdermal microneedle sensor is demonstrated for potassium by integrating a hollow microneedle with a microfluidic chip to extract fluid through a channel towards a downstream solid-state ion-selective-electrode (ISE). 3D porous carbon and 3D porous graphene electrodes, made via interference lithography, are compared as solid-state transducers for ISE's and evaluated for electrochemical performance, stability, and selectivity. The porous carbon K(+) ISE's show better performance than the porous graphene K(+) ISE's, capable of measuring potassium across normal physiological concentrations in the presence of interfering ions with greater stability. This new microfluidic/microneedle platform shows promise for medical applications.

Medical subject headings