The properties of an improvised piston pump for the rapid delivery of intravenous fluids.
other · Level V
Where this comes from
- Record sourced from PubMed, PMID 24383681.
- Also identified by DOI 10.1111/anae.12463.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To maximise the effect of a small fluid load, it is occasionally desirable to bolus manually with multiple depressions of a large-capacity syringe. This is usually achieved by placing the syringe on the side port of a three-way tap. We modified this technique by placing two-one-way valves in line with the three-way tap, effectively creating a piston pump, the infusion rates via which we compared with those achieved by an inflatable pressure-infuser in a simulated resuscitation. Fluid flow was faster using the piston pump than with the pressure-infuser (mean (SD) time to infuse 2000 ml saline 0.9% via a 16-G cannula 352 (10) s vs 495 (19) s, respectively, p < 0.0001). The piston pump appears to have potential for both tight control of fluid delivery and major high-volume resuscitation. The lightweight nature of the pump and its lack of reliance on gravity may also make it suitable for the pre-hospital setting.
Medical subject headings
- Fluid Therapy
- Infusion Pumps