Quantitative analysis of coronary vessels with optimized intracoronary CT number.
Where this comes from
- Record sourced from PubMed, PMID 24409326.
- Also identified by DOI 10.1371/journal.pone.0085312 and PMC identifier 3883697.
- Licence recorded as CC BY.
- The licence permits redistribution, so the abstract is shown in full and the full text is available from the publisher.
Abstract
Variability in intracoronary computed tomography (CT) number may influence vessel quantification. We confirmed the feasibility of a novel method for measuring vessel diameter and area using coronary CT angiography (CCTA) with an optimized intracoronary CT number, 350 HU. We performed intravascular ultrasound (IVUS) imaging in 52 patients with significant stenosis detected by coronary CT angiography targeting 350 HU using a CT number-controlling system. We measured 0-to-0 HU distances in the cross-sectional coronary images of 32 patients. We analyzed the ratio of 0-to-0 HU distances in CT images to media-to-media distances in IVUS images (C:I ratio). The area of ≥0 HU for 103 representative points in the remaining 20 patients was compared to the area of the traced external elastic membrane (EEM) in IVUS images. There was a strong correlation between 0-to-0 HU distance in CT images and media-to-media diameter in IVUS images (r = 0.97, p<0.001). The C:I ratio was 1.1. EEM area was estimated by dividing the area of ≥0 HU by the square of C:I. There was also a strong correlation between the estimated EEM area and the EEM area in IVUS images (r = 0.95, p<0.001). Media-to-media diameter and EEM area can be estimated by CCTA targeting the optimized intracoronary CT number when blood vessel borders are defined at 0 HU.
Medical subject headings
- Coronary Angiography
- Coronary Vessels
- Ultrasonography, Interventional