Failure of a carbon fiber-reinforced polymer implant used for transforaminal lumbar interbody fusion.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 24436878.
- Also identified by DOI 10.1055/s-0033-1343075 and PMC identifier 3854590.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Lumbar interbody fusion is a common procedure owing to the high prevalence of degenerative spinal disorders. During such procedures, carbon fiber-reinforced polymer (CFRP) cages are frequently utilized to fill the void created between adjacent vertebral bodies, to provide mechanical stability, and to carry graft material. Failure of such implants can lead to significant morbidity. We discuss the possible causes leading to the failure of a CFRP cage in a patient with rheumatoid arthritis. Review of a 49-year-old woman who underwent revision anterior lumbar interbody fusion 2 years after posterior instrumentation and transforaminal lumbar interbody fusion at L4-L5 and L5-S1. The patient developed pseudarthrosis at the two previously fused levels with failure of the posterior instrumentation. Revision surgery reveled failure with fragmentation of the CFRP cage at the L5-S1 level. CFRP implants can break if mechanical instability or nonunion occurs in the spinal segments, thus emphasizing the need for optimizing medical management and meticulous surgical technique in achieving stability.
Anatomy
- lumbar spine