Robot-assisted total knee arthroplasty accurately restores the joint line and mechanical axis. A prospective randomised study.

Liow, Ming Han Lincoln; Xia, Zhan; Wong, Merng Koon; Tay, Keng Jin; Yeo, Seng Jin; Chin, Pak Lin · J Arthroplasty · 2014

rct · Level II

Where this comes from

Abstract

Robot-assisted Total Knee Arthroplasty (TKA) improves the accuracy and precision of component implantation and mechanical axis (MA) alignment. Joint-line restoration in robot-assisted TKA is not widely described and joint-line deviation of>5mm results in mid-flexion instability and poor outcomes. We prospectively randomised 60 patients into two groups: 31 patients (robot-assisted), 29 patients (conventional). No MA outliers (>±3° from neutral) or notching was noted in the robot-assisted group as compared with 19.4% (P=0.049) and 10.3% (P=0.238) respectively in the conventional group. The robot-assisted group had 3.23% joint-line outliers (>5mm) as compared to 20.6% in the conventional group (P=0.049). Robot-assisted TKA produces similar short-term clinical outcomes when compared to conventional methods with reduction of MA alignment and joint-line deviation outliers.

Medical subject headings

Anatomy