Pulse lavage is inadequate at removal of biofilm from the surface of total knee arthroplasty materials.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 24439797.
- Also identified by DOI 10.1016/j.arth.2013.12.012.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
In acute periprosthetic infection, irrigation and debridement with component retention has a high failure rate in some studies. We hypothesize that pulse lavage irrigation is ineffective at removing biofilm from total knee arthroplasty (TKA) components. Staphylococcus aureus biofilm mass and location was directly visualized on arthroplasty materials with a photon collection camera and laser scanning confocal microscopy. There was a substantial reduction in biofilm signal intensity, but the reduction was less than a ten-fold decrease. This suggests that irrigation needs to be further improved for the removal of biofilm mass below the necessary bioburden level to prevent recurrence of acute infection in total knee arthroplasty.
Medical subject headings
- Arthroplasty, Replacement, Knee
- Biofilms
- Equipment Contamination
- Knee Prosthesis
- Staphylococcus aureus
- Therapeutic Irrigation
Anatomy
- knee