Cross-validation of two commercial methods for volumetric high-resolution dose reconstruction on a phantom for non-coplanar VMAT beams.
other · Level V
Where this comes from
- Record sourced from PubMed, PMID 24440044.
- Also identified by DOI 10.1016/j.radonc.2013.12.011.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Delta(4) (ScandiDos AB, Uppsala, Sweden) and ArcCHECK with 3DVH software (Sun Nuclear Corp., Melbourne, FL, USA) are commercial quasi-three-dimensional diode dosimetry arrays capable of volumetric measurement-guided dose reconstruction. A method to reconstruct dose for non-coplanar VMAT beams with 3DVH is described. The Delta(4) 3D dose reconstruction on its own phantom for VMAT delivery has not been thoroughly evaluated previously, and we do so by comparison with 3DVH. Reconstructed volumetric doses for VMAT plans delivered with different table angles were compared between the Delta(4) and 3DVH using gamma analysis. The average γ (2% local dose-error normalization/2mm) passing rate comparing the directly measured Delta(4) diode dose with 3DVH was 98.2 ± 1.6% (1SD). The average passing rate for the full volumetric comparison of the reconstructed doses on a homogeneous cylindrical phantom was 95.6 ± 1.5%. No dependence on the table angle was observed. Modified 3DVH algorithm is capable of 3D VMAT dose reconstruction on an arbitrary volume for the full range of table angles. Our comparison results between different dosimeters make a compelling case for the use of electronic arrays with high-resolution 3D dose reconstruction as primary means of evaluating spatial dose distributions during IMRT/VMAT verification.
Medical subject headings
- Phantoms, Imaging
- Radiotherapy Dosage
- Radiotherapy, Intensity-Modulated