Numerical simulation of arterial dissection during balloon angioplasty of atherosclerotic coronary arteries.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 24480707.
- Also identified by DOI 10.1016/j.jbiomech.2014.01.009 and PMC identifier 3947651.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Balloon angioplasty is a standard clinical treatment for symptomatic coronary artery disease. In this procedure, controlled damage is applied intraluminally to the wall of a stenotic artery. Dissection of the coronary artery is a commonly observed clinical complication of angioplasty; however, not all dissections can be detected angioscopically. This work focuses on studying the dissection mechanisms triggered during the early stages of angioplasty in an atherosclerotic coronary artery, addressing the problem by means of a parametric study based on a simplified finite element model and cohesive interface modeling. Our results emphasize the presence of several damage mechanisms, at different locations, that are triggered near the very beginning of the process and evolve competitively, depending on both geometry and material properties of the atherosclerotic vessel. Small-scale damage was evidenced, which would not be detectable by angiography or intravascular ultrasound, but could potentially be sufficient to stimulate smooth muscle cell activation, promoting late-onset complications such as restenosis.
Medical subject headings
- Aortic Dissection
- Angioplasty, Balloon, Coronary
- Coronary Aneurysm
- Coronary Artery Disease
- Finite Element Analysis
- Models, Cardiovascular