Disassembly of threaded junction between stem extension and femoral component in a total stabilizer revision total knee arthroplasty.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 24508094.
- Also identified by DOI 10.1016/j.knee.2014.01.004.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We report an unusual case of disassembly of threaded junction between stem extension and femoral component in a 6-year-old total stabilizer revision total knee arthroplasty. Aseptic loosening of femur component will tend to cause increased stresses at the stem-condylar junction of a modular femoral prosthesis, especially if femur component is flexed. The modular junction may therefore be unable to withstand the concentrated force and lead to disassembly.