Retrograde trans-anterior communicating artery rescue of unopened Pipeline Embolization Device with balloon dilation: complication management.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 24510421.
- Also identified by DOI 10.1136/neurintsurg-2013-011009.rep.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
As the use of the Pipeline Embolization Device (PED) for the treatment of complex intracranial aneurysms rises, knowledge about complications continues to accumulate amidt a paucity of reports on techniques and rescue strategies. We describe the case of a 70-year-old woman who presented with worsening reto-orbital left-sided pain and a large cavernous aneurysm. The patient underwent endovascular treatment with PED, and there was difficulty delivering the device due to significant vascular tortuosity. This resulted in poor PED deployment as the proximal end failed to open. Increasingly aggressive strategies were attempted to open the device, which resulted in an iatrogenic carotid cavernous fistula. We were finally able to rescue the device and open its proximal end with balloon inflation after using a contralateral trans-anterior communicating artery approach and crossing the PED in a retrograde fashion. Excessive vascular tortuosity poses a genuine risk of PED malfunction and poor deployment. Although we were able to rescue the device and our patient had no permanent morbidity, difficult vascular anatomy rendered the procedure extremely complicated with dreaded complications.
Medical subject headings
- Angioplasty, Balloon
- Anterior Cerebral Artery
- Embolization, Therapeutic