A pitfall during endobronchial ultrasound-guided transbronchial forceps biopsy of the mediastinal lymph nodes.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 24580955.
- Also identified by DOI 10.1016/j.athoracsur.2013.11.061.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The high diagnostic yield and favorable safety profile of endobronchial ultrasound-guided transbronchial forceps biopsy of the mediastinal lymph nodes have been recently demonstrated. We report an unusual technical problem during endobronchial ultrasound-guided transbronchial forceps biopsy that could be a prerequisite for severe complications. A rupture of the steering band precluded closure of the forceps jaws opened in the subcarinal lymph node. A solution to the problem is presented, together with other procedure-related complications reported in the literature. The report emphasizes that a dysfunction of the forceps steering band can result in severe complications when it occurs during transbronchial sampling of mediastinal lesions.
Medical subject headings
- Endoscopic Ultrasound-Guided Fine Needle Aspiration
- Lymph Nodes