Load sharing and ligament strains in balanced, overstuffed and understuffed UKA. A validated finite element analysis.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 24581895.
- Also identified by DOI 10.1016/j.arth.2014.01.020.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The aim of this study was to quantify the effects of understuffing and overstuffing UKA on bone stresses, load distribution and ligament strains. For that purpose, a numerical knee model of a cadaveric knee was developed and was validated against experimental measurements on that same knee. Good agreement was found among the numerical and experimental results. This study showed that, even if a medial UKA is well-aligned with normal soft tissue tension and with correct thickness of the tibia component, it induces a stiffness modification in the joint that alters the load distribution between the medial and lateral compartments, the bone stress and the ligament strain potentially leading to an osteoarthritic progression.
Medical subject headings
- Arthroplasty, Replacement, Knee
- Knee Joint
- Medial Collateral Ligament, Knee
Anatomy
- knee
- tibia