Comparison of real-time water proton spectroscopy and echo-planar imaging sensitivity to the BOLD effect at 3 T and at 7 T.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 24614912.
- Also identified by DOI 10.1371/journal.pone.0091620 and PMC identifier 3948886.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Gradient-echo echo-planar imaging (GE EPI) is the most commonly used approach to assess localized blood oxygen level dependent (BOLD) signal changes in real-time. Alternatively, real-time spin-echo single-voxel spectroscopy (SE SVS) has recently been introduced for spatially specific BOLD neurofeedback at 3 T and at 7 T. However, currently it is not known how neurofeedback based on real-time SE SVS compares to real-time GE EPI-based. We therefore compared both methods at high (3 T) and at ultra-high (7 T) magnetic field strengths. We evaluated standard quality measures of both methods for signals originating from the motor cortex, the visual cortex, and for a neurofeedback condition. At 3 T, the data quality of the real-time SE SVS and GE EPI R2* estimates were comparable. At 7 T, the data quality of the real-time GE EPI acquisitions was superior compared to those of the real-time SE SVS. Despite the somehow lower data quality of real-time SE SVS compared to GE EPI at 7 T, SE SVS acquisitions might still be an interesting alternative. Real-time SE SVS allows for a direct and subject-specific T2* estimation and thus for a physiologically more plausible neurofeedback signal.
Medical subject headings
- Computer Systems
- Echo-Planar Imaging
- Oxygen
- Protons
- Spectrum Analysis
- Water