Electrosurgery: part II. Technology, applications, and safety of electrosurgical devices.

Taheri, Arash; Mansoori, Parisa; Sandoval, Laura F; Feldman, Steven R; Pearce, Daniel; Williford, Phillip M · J Am Acad Dermatol · 2014

review · Level V

Where this comes from

Abstract

Electrosurgical currents can be delivered to tissue in monopolar or bipolar and monoterminal or biterminal modes, with the primary difference between these modes being their safety profiles. A monopolar electrosurgical circuit includes an active electrode and a dispersive (return) electrode, while there are 2 active electrodes in bipolar mode. In monoterminal mode, there is an active electrode, but there is no dispersive electrode connected to the patient's body and instead the earth acts as the return electrode. Biterminal mode uses a dispersive electrode connected to the patient's body, has a higher maximum power, and can be safer than monoterminal mode in certain situations. Electrosurgical units have different technologies for controlling the output power and for providing safety. A thorough understanding of these technologies helps with a better selection of the appropriate surgical generator and modes.