Intraocular lens injector-induced stress on the corneal incisions during lens implantation.

Kleinmann, Guy; Kleinmann, Itzhak · Am J Ophthalmol · 2014

biomechanical · Level V

Where this comes from

Abstract

To describe a model for calculating induced stress on corneal incisions during intraocular lens implantation, and to compare the nozzles of common injectors using 2 incision sizes. Experimental study. A finite element calculation model was developed and used to compare 7 commercially available injectors in widespread clinical use. The injectors' characteristics were measured and correlated for the stress they induced for 2.4- or 2.2-mm corneal incisions. Each injector created a different level of stress on the corneal incision. The stress was highest at the incision margins, and its level correlated with the injector's external circumference. The induced stress on 2.2-mm incision margins was about 9% higher compared to the induced stress on 2.4-mm incision margins. A model for comparing stress induced on corneal incisions during intraocular lens implantation by 7 injectors revealed different levels of induced stress on the incision margins. It is therefore recommended that the choice of injector be matched to the size of the corneal incision.

Medical subject headings