Three-dimensional, sharp-tipped electrodes concentrate applied fields to enable direct electrical release of intact biomarkers from cells.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 24695906.
- Also identified by DOI 10.1039/c4lc00144c.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Biomarkers such as proteins and nucleic acids released from human cells, bacteria, and viruses offer a wealth of information pertinent to diagnosis and treatment ranging from cancer to infectious disease. The release of these molecules from within cells is a crucial step in biomarker analysis. Here we show that purely electric-field-driven lysis can be achieved, inline, within a microfluidic channel; that it can produce highly efficient lysis and biomarker release; and, further, that it can do so with minimal degradation of the released biomarkers. Central to this new technology is the use of three-dimensional sharp-tipped electrodes (3DSTEs) in lysis, which we prove using experiment and finite-element modeling produce the electric field concentration necessary for efficient cell wall rupture.
Medical subject headings
- Biomarkers
- Microfluidic Analytical Techniques