An in vitro evaluation of the impact of eccentric deployment on transcatheter aortic valve hemodynamics.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 24719050.
- Also identified by DOI 10.1007/s10439-014-1008-6.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Patients with aortic stenosis present with calcium deposits on the native aortic valve, which can result in non-concentric expansion of Transcatheter Aortic Valve Replacement (TAVR) stents. The objective of this study is to evaluate whether eccentric deployment of TAVRs lead to turbulent blood flow and blood cell damage. Particle Image Velocimetry was used to quantitatively characterize fluid velocity fields, shear stress and turbulent kinetic energy downstream of TAVRs deployed in circular and eccentric orifices representative of deployed TAVRs in vivo. Effective orifice area (EOA) and mean transvalvular pressure gradient (TVG) values did not differ substantially in circular and eccentric deployed valves, with only a minor decrease in EOA observed in the eccentric valve (2.0 cm(2) for circular, 1.9 cm(2) for eccentric). Eccentric deployed TAVR lead to asymmetric systolic jet formation, with increased shear stresses (circular = 97 N/m(2) vs. eccentric = 119 N/m(2)) and regions of turbulence intensity (circular = 180 N/m(2) vs. eccentric = 230 N/m(2)) downstream that was not present in the circular deployed TAVR. The results of this study indicate that eccentric deployment of TAVRs can lead to altered flow characteristics and may potentially increase the hemolytic potential of the valve, which were not captured through hemodynamic evaluation alone.
Medical subject headings
- Aortic Valve
- Cardiac Catheterization
- Heart Valve Prosthesis Implantation
- Models, Cardiovascular
- Pulsatile Flow