Component size mismatch of metal on metal hip arthroplasty: an avoidable never event.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 24726175.
- Also identified by DOI 10.1016/j.arth.2014.03.008.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Size mismatch of components used in total hip arthroplasty is a serious, preventable patient safety incident of unknown prevalence as many cases are not detected. Component size mismatch was found in 11 cases (0.9%) at our retrieval centre. All cases of mismatch were not detected on plain radiograph during routine clinical follow up and blood metal ion levels were elevated above the MHRA action level of 7 ppb. Root cause analysis identified manufacturer, hospital and surgeon factors that need to be addressed to reduce the incidence of this avoidable clinical problem. Retrieval analysis is the only method of confirming size mismatch and is likely to be under-represented in National Joint Registries that record the indication for revision at the time of revision.
Medical subject headings
- Arthroplasty, Replacement, Hip
- Hip Prosthesis
- Metal-on-Metal Joint Prostheses
- Patient Safety
- Prosthesis Failure
Anatomy
- hip