Assessment of accuracy of marker ball placement in pre-operative templating for total hip arthroplasty.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 24736293.
- Also identified by DOI 10.1016/j.arth.2014.03.013.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We report the accuracy of positioning of the calibration ball in the process of pre-operative templating for total hip arthroplasty (THA). The ball should be placed in the coronal plane of the hip to provide suitable accuracy. We reviewed 112 post-operative THA radiographs where a calibration ball had been placed. We templated the femoral head size of the implant after calibrating the templating system from the ball. The calibrated femoral head diameter was compared to the known prosthetic head size. A percentage error was calculated. Overall, incorrect placement of the calibration ball resulted in a mean percentage error in templating of 6.8% (median 5.7%; range 0-26%). Such error carries implications with the templating process and may result in incorrect component sizes, leg lengths and offset.