Detection of lung tumors in mice using a 1-tesla compact magnetic resonance imaging system.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 24743153.
- Also identified by DOI 10.1371/journal.pone.0094945 and PMC identifier 3990561.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Due to their small size, lung tumors in rodents are typically investigated using high-field magnetic resonance (MR) systems (4.7 T or higher) to achieve higher signal-to-noise ratios, although low-field MR systems are less sensitive to susceptibility artifacts caused by air in the lung. We investigated the feasibility of detecting lung tumors in living, freely breathing mice with a 1-T compact permanent magnet MR system. In total, 4 mice were used, and MR images of mouse lungs were acquired using a T1-weighted three-dimensional fast low-angle shot sequence without cardiac or respiratory gating. The delineation and size of lung tumors were assessed and compared with histopathological findings. Submillimeter lesions were demonstrated as hyperintense, relative to the surrounding lung parenchyma, and were delineated clearly. Among the 13 lesions validated in histopathological sections, 11 were detected in MR images; the MR detection rate was thus 84.6%. A strong correlation was obtained in size measurements between MR images and histological sections. Thus, a dedicated low-field MR system can be used to detect lung tumors in living mice noninvasively without gating.
Medical subject headings
- Lung Neoplasms
- Magnetic Resonance Imaging
- Neoplasms, Experimental