In vivo loss of cement-bone interlock reduces fixation strength in total knee arthroplasties.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 24777486.
- Also identified by DOI 10.1002/jor.22634 and PMC identifier 4074492.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Prevention of aseptic loosening of total knee arthroplasties (TKAs) remains a clinical challenge. Understanding how changes in morphology at the implant-bone interface with in vivo service affect implant stability and strength could lead to new approaches to mitigate loosening. Enbloc TKA retrievals and freshly-cemented TKA tibial components were used to determine if the mechanical strength of the interface depended on the amount of cement-bone interlock and the morphology of the supporting bone under the cement layer. Implants were sectioned into small specimens of the cement-interface-bone from under the tibial tray. Micro-CT scans were used to document interlock morphology and architecture of the supporting trabecular bone. Axial compression tests were used to assess mechanical behavior. Postmortem retrievals had lower contact fraction (42 ± 55%) compared to freshly-cemented constructs (121 ± 61%) (p = 0.0008). Supporting bone architecture parameters were not different for the two groups. Increased interface contact fraction and supporting bone volume fraction (BV/TV) were positive predictors of interface strength (r(2) = 0.72, p = 0.0001). For the same supporting bone BV/TV, postmortem specimens had weaker interfaces; they were also more compliant. Cemented TKAs with in vivo service experience a loss of fixation strength and increased micro-motion due to the loss of cement-bone interlock.
Medical subject headings
- Arthroplasty, Replacement, Knee
- Bone Cements
- Equipment Failure Analysis
- Knee Prosthesis
Anatomy
- knee