Conformity and robustness of gated rescanned carbon ion pencil beam scanning of liver tumors at NIRS.

Mori, Shinichiro; Zenklusen, Silvan; Inaniwa, Taku; Furukawa, Takuji; Imada, Hiroshi; Shirai, Toshiyuki; Noda, Koji; Yasuda, Shigeo · Radiother Oncol · 2014

basic_science · Level V

Where this comes from

Abstract

Pencil beam scanning offers excellent conformity, but is sensitive to organ motion. We conducted a simulation study to validate our rescanning approach in combination with gating in the irradiation of liver tumors. 4DCT imaging was performed under free-breathing conditions in 30 patients with hepatocellular carcinoma. Dose distributions for a two-field approach were calculated for layered phase controlled rescannings (PCR) under organ motion conditions. A total dose of 45 Gy(RBE) was delivered to respective field-specific target volumes (FTVs) in two fractions, each composed of two orthogonal uniform fields of 11.25 Gy(RBE) at beam angles of either 0° and 90° or 0° and 270°. The number of rescannings was changed from 1 to 10. Good dose conformity was achieved with 4× PCR or more, and over 95% of the prescribed dose was delivered to the CTV independent of the use of gating. D95, Dmax/min and dose homogeneity were similar with or without gating, whereas V10 dose to the liver as well as maximal doses to healthy tissue (esophagus and cord) were about 40% lower with gating. However, total time increased by about 50% with gating. Gated rescanning provides good target coverage and homogeneity with maximal sparing of healthy tissue. Our results suggest that carbon-ion pencil beam scanning may soon be available for the safe treatment of liver tumors.