Paclitaxel-coated balloons and aneurysm formation in peripheral vessels.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 24801552.
- Also identified by DOI 10.1016/j.jvs.2014.03.291.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We report two cases of early aneurysmal vessel dilatation after a paclitaxel-coated balloon (PCB) was used for angioplasty of the peripheral vessels. The first case refers to a failing vein bypass with a tight proximal anastomotic stenosis, whereas the second refers to a distal tibial artery occlusion. A PCB was used to treat both patients. Aneurysmal dilatation of the previously treated segment was noted in both patients during subsequent follow-up imaging. In the absence of other causal factors, we attribute both cases to PCB application. The aneurysms that formed had no detrimental effect on the patients' health and required no further treatment; however, it is important to bear in mind this potential risk of presumed paclitaxel toxicity.