A prospective randomised controlled trial of the LMA Supreme vs cuffed tracheal tube as the airway device during percutaneous tracheostomy.
rct · Level II
Where this comes from
- Record sourced from PubMed, PMID 24803133.
- Also identified by DOI 10.1111/anae.12595.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We studied the performance of the LMA Supreme against a cuffed tracheal tube, our standard method of airway control during percutaneous tracheostomy, in 50 consecutive patients from three general critical care units. The primary outcome measure was adequacy of ventilation calculated as the difference in arterial carbon dioxide tension before and after tracheostomy. On an intention-to-treat analysis, there was no difference in the increase in arterial carbon dioxide tension between groups, with a median (IQR [range]) for the LMA Supreme of 0.9 (0.3-1.6 [0-2.8]) kPa, and for the tracheal tube of 0.8 (0.4-1.2 [0-2.5]) kPa, p = 0.82. Eight patients out of 25 (32%) crossed over from the LMA Supreme group to the tracheal tube group before commencement of tracheostomy due to airway or ventilation problems, compared with none out of 25 in the tracheal tube group, p = 0.01, and tracheostomy was postponed in two patients in the LMA Supreme group due to poor oxygenation. There were more clinically important complications in the LMA Supreme group compared with the tracheal tube group.
Medical subject headings
- Intubation, Intratracheal
- Respiration, Artificial
- Tracheostomy