Image quality and radiation dose of low tube voltage 3rd generation dual-source coronary CT angiography in obese patients: a phantom study.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 24816937.
- Also identified by DOI 10.1007/s00330-014-3194-x.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To assess the influence of tube potential on radiation dose and image quality of third-generation dual-source coronary CT angiography (CTA) in a phantom simulating an obese patient. A thoracic phantom was equipped with tubular inserts containing iodine solution and water. A soft-tissue-equivalent ring around the phantom simulated an obese patient. Images were acquired at tube potentials of 80, 100, 120 and 140 kV with second-generation dual-source CT (DSCT) and 70-150 kV (in 10-kV increments) with third-generation DSCT. Contrast-to-noise ratio (CNR) was calculated and CT dose index was recorded. With second-generation DSCT, CNR was highest for 120 kV (19.0) and decreased with lower tube potential (12.0 at 80 kV) owing to disproportionately increased image noise. With third-generation DSCT, 70- and 80-kV acquisitions showed a smaller increase in noise. CNRs for third-generation DSCT were highest for 70 and 80 kV (21.1 and 21.2, respectively). Compared to 120 kV, radiation dose was 68% and 49% lower at 70 kV and 80 kV, respectively. Third-generation DSCT enables one to perform coronary CTA at 70-80 kV in obese patients without compromising CNR and thus reduces radiation dose by 49-68%. • Low tube potential CT angiography is currently not suitable for obese patients. • Third-generation DSCT offers substantially increased tube power at low tube potential. • This enables one to perform coronary CT angiography at 70-80 kV in obese patients. • Signal-to-noise ratio is maintained owing to increased tube current. • This approach can be expected to reduce radiation dose by 49-68%.
Medical subject headings
- Coronary Angiography
- Coronary Artery Disease
- Image Processing, Computer-Assisted
- Multidetector Computed Tomography
- Obesity
- Phantoms, Imaging