How does laxity after single radius total knee arthroplasty compare with the native knee?
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 24841798.
- Also identified by DOI 10.1002/jor.22645.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Patients with total knee arthroplasties (TKAs) continue to report dissatisfaction in functional outcome. Stability is a major factor contributing to functionality of TKAs. Implants with single-radius (SR) femoral components are proposed to increase stability throughout the arc of flexion. Using computer navigation and loaded cadaveric legs, we characterized the "envelope of laxity" (EoL) offered by a SR cruciate retaining (CR)-TKA compared with that of the native knee through the arc of flexion in terms of anterior drawer, varus/valgus stress, and internal/external rotation. In both the native knee and the TKA laxity increased with increasing knee flexion. Laxities measured in the three planes of motion were generally comparable between the native knee and TKA from 0° to 110° of flexion. Our results indicate that the SR CR-TKA offers appropriate stability in the absence of soft tissue deficiency.