Optoacoustic imaging of subcutaneous microvasculature with a class one laser.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 24876111.
- Also identified by DOI 10.1109/TMI.2014.2326552.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We developed a combined imaging platform allowing optoacoustic and ultrasound imaging based on a low energy laser and a handheld probe. The device is based on a sensitive single element 35-MHz focused transducer, a 2-D piezoscanner and a dual-wavelength switchable Nd:YAG laser. Acoustical detection and optical illumination are confocal for optimization of optoacoustic signal-to-noise ratio. The system allows to scan over a range up to 12 mm ×12 mm in xy-direction with an isotropic lateral resolution of about 90 μm. Although the device is a class 1 laser product having pulse energies in the range, in vivo images of subcutaneous microvasculature could be obtained from human skin with signal-to-noise levels as good as 20 dB.
Medical subject headings
- Diagnostic Imaging
- Image Processing, Computer-Assisted
- Microvessels
- Photoacoustic Techniques