Two-dimensional compressed sensing using the cross-sampling approach for low-field MRI systems.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 24879645.
- Also identified by DOI 10.1109/TMI.2014.2326864.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
A compressed sensing method using a cross sampling and self-calibrated off-resonance correction is proposed. Estimation of the magnetic field inhomogeneity based on image registration enables the off-resonance correction with no additional radio-frequency pulses or acquisitions. In addition to this advantage, a fast and straightforward calculation was achieved by using the first-order components of the magnetic field inhomogeneity. Imaging experiments using a phantom and a chemically fixed mouse demonstrated practical benefits in improving blurring and artifacts in magnetic resonance images in low field magnetic resonance imaging systems.
Medical subject headings
- Image Processing, Computer-Assisted
- Magnetic Resonance Imaging