A review of safety, quality management, and practice guidelines for high-dose-rate brachytherapy: executive summary.
review · Level V
Where this comes from
- Record sourced from PubMed, PMID 24890345.
- Also identified by DOI 10.1016/j.prro.2013.12.005.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
This white paper was commissioned by the American Society for Radiation Oncology (ASTRO) Board of Directors to evaluate the status of safety and practice guidance for high-dose-rate (HDR) brachytherapy. Given the maturity of HDR brachytherapy technology, this white paper considers, from a safety point of view, the adequacy of general physics and quality assurance guidance, as well as clinical guidance documents available for the most common treatment sites. The rate of medical events in HDR brachytherapy procedures in the United States in 2009 and 2010 was 0.02%, corresponding to 5-sigma performance. The events were not due to lack of guidance documents but failures to follow those recommendations or human failures in the performance of tasks. The white paper summarized by this Executive Summary reviews current guidance documents and offers recommendations regarding their application to delivery of HDR brachytherapy. It also suggests topics where additional research and guidance is needed.
Medical subject headings
- Brachytherapy
- Patient Safety
- Practice Guidelines as Topic
- Radiation Oncology