Midterm results of a femoral stem with a modular neck design: clinical outcomes and metal ion analysis.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 24895065.
- Also identified by DOI 10.1016/j.arth.2014.04.039.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Modular neck femoral stems have a higher-than-anticipated rate of failure in registry results, but large single-center cohort studies are lacking. This is a retrospective cohort of 152 hips implanted with a single titanium stem with a modular titanium neck, presenting clinical, radiographic, and metal ion results at a mean 4.5-year follow-up. Five hips were revised during the study period, for an overall Kaplan-Meier survival of 0.894 at 8 years. There was one modular neck fracture (0.66%), but others demonstrated corrosion or adverse tissue reaction. Serum metal levels demonstrated wide variability. Despite good clinical results in the majority of patients, we confirmed an increased rate of femoral revision at mid-term follow-up, and therefore urge caution in the use of this particular stem design.
Medical subject headings
- Arthroplasty, Replacement, Hip
- Hip Prosthesis
- Osteoarthritis, Hip
- Prosthesis Failure
- Titanium
Anatomy
- femur
- hip