Monolithically integrated mid-infrared lab-on-a-chip using plasmonics and quantum cascade structures.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 24905443.
- Also identified by DOI 10.1038/ncomms5085 and PMC identifier 4059934.
- Licence recorded as CC BY-NC-SA.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The increasing demand of rapid sensing and diagnosis in remote areas requires the development of compact and cost-effective mid-infrared sensing devices. So far, all miniaturization concepts have been demonstrated with discrete optical components. Here we present a monolithically integrated sensor based on mid-infrared absorption spectroscopy. A bi-functional quantum cascade laser/detector is used, where, by changing the applied bias, the device switches between laser and detector operation. The interaction with chemicals in a liquid is resolved via a dielectric-loaded surface plasmon polariton waveguide. The thin dielectric layer enhances the confinement and enables efficient end-fire coupling from and to the laser and detector. The unamplified detector signal shows a slope of 1.8-7 μV per p.p.m., which demonstrates the capability to reach p.p.m. accuracy over a wide range of concentrations (0-60%). Without any hybrid integration or subwavelength patterning, our approach allows a straightforward and cost-saving fabrication.
Medical subject headings
- Equipment Design
- Miniaturization
- Spectroscopy, Fourier Transform Infrared
- Surface Plasmon Resonance