Accuracy of MR in growth plate measurement.

Shiguetomi-Medina, Juan Manuel; Rahbek, Ole; Ringgaard, Steffen; Kristiansen, Maja Sofie; Stødkilde-Jørgensen, Hans; Møller-Madsen, Bjarne · Skeletal Radiol · 2014

basic_science · Level V

Where this comes from

Abstract

To analyze the accuracy of growth-plate thickness measurements detected on 1.5-T and 7-T MR images using histology sections as a standard of reference. Four defrosted pig tibiae were 1.5-T MR scanned and one fresh tibia was 7-T MR scanned. The height of the growth plate was measured and compared to histology. Histology measurements showed a mean growth plate thickness of 467 μm (SD = 82.2). The mean growth plate thickness measured in the 7-T MR images was 465 μm (SD = 62.2) and 1,325 μm (SD = 183.5) on 1.5-T MR measurements. We found a better correspondence between the growth plate thickness measured on the 7-T MR and histology samples compared to 1.5 T. The growth plate can be identified and measured with high accuracy using 7-T MR. 1.5-T MR can only describe some morphological characteristics.

Medical subject headings

Anatomy