Clinical implementation of MRI of joint arthroplasty.

Koff, Matthew F; Shah, Parina; Potter, Hollis G · AJR Am J Roentgenol · 2014

review · Level V

Where this comes from

Abstract

The purpose of this article is to explain the basic physics of imaging patients with metal implants, explain conflicting information regarding MRI scanning of "MR Conditional" devices, and relate our experience of scanning total joint arthroplasty (TJA) at our institution. MRI near TJA is effective with appropriate imaging protocols and standardized safety precautions. Strict adherence to MR Conditional labeling may preclude broad use of MRI for TJA assessment.

Medical subject headings