Which spring is the best? Comparison of methods for virtual stenting.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 24956618.
- Also identified by DOI 10.1109/TBME.2014.2311856.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
This paper presents a methodology for modeling the deployment of implantable devices used in minimally invasive vascular interventions. Motivated by the clinical need to perform preinterventional rehearsals of a stent deployment, we have developed methods enabling virtual device placement inside arteries, under the constraint of real-time application. This requirement of rapid execution narrowed down the search for a suitable method to the concept of a dynamic mesh. Inspired by the idea of a mesh of springs, we have found a novel way to apply it to stent modeling. The experiments conducted in this paper investigate properties of the stent models based on three different spring types: lineal, semitorsional, and torsional springs. Furthermore, this paper compares the results of various deployment scenarios for two different classes of devices: a stent graft and a flow diverter. The presented results can be of a high-potential clinical value, enabling the predictive evaluation of the outcome of a stent deployment treatment.
Medical subject headings
- Blood Vessel Prosthesis
- Blood Vessel Prosthesis Implantation
- Endovascular Procedures
- Minimally Invasive Surgical Procedures
- Models, Cardiovascular
- Stents