Incidence of patellar clunk syndrome in fixed versus high-flex mobile bearing posterior-stabilized total knee arthroplasty.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 24961894.
- Also identified by DOI 10.1016/j.arth.2014.05.011.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The geometry of the intercondylar box plays a significant role in the development of patellar clunk syndrome. We reviewed the incidence of patella clunk at mid-to-long-term follow-up of a rotating high-flex versus fixed bearing posterior stabilized TKA design. 188-mobile and 223-fixed bearing TKAs were reviewed for complications, incidence of patellar clunk, treatment, recurrence rates, range of motion, and patient satisfaction. Patellar clunk developed in 22 knees in the mobile (11.7%) and in 4 (1.8%) in the fixed bearing group (P<0.001). 23 out of 26 cases resolved with a single arthroscopic treatment and 2 resolved with a second procedure. The mean postoperative range of motion was 122.4°. All but one patient reported overall satisfaction with the index procedure. In contrast with other recent studies we found a significant incidence of patellar clunk in high-flex mobile bearings. Despite the high rate of patellar clunk syndrome, overall patients did well and were satisfied with their outcomes.
Medical subject headings
- Arthroplasty, Replacement, Knee
- Joint Diseases
- Knee Prosthesis
Anatomy
- knee